6.1 Electrocardiography (ECG)
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 69
Menu option Menu screen
HR/PR beat volume
– volume of the audible
tone that the Tempus emits every time a
heartbeat is detected. Use the up and down
arrows to adjust the volume from zero (no
tone) to 100%. The tone is based on the
HR/PR source setting (see above): when set
to ECG, QRS beats trigger a single tone at
1 kHz. When set to Pulse, pulse beats trigger
pulse tones – see “6.3.6 Pulse oximeter
settings”.
Pacemaker indication – turning this on will
show a vertical dotted line on the ECG
waveform when the ECG is connected to a
patient wearing a pacemaker.
Monitoring filter – use the arrows to select a
filtering option.
12 lead filter – use the arrows to select a
filtering option.
ECG Power Filter – this is set to ON by
default, switch it off if you wish the ECG to be
presented without the Power noise being
filtered out.
ST lead selection screen
Note
The ECG waveforms displayed on the home screen will change to match the selected ST
leads.
ST Lead Selection
Then select the
lead you wish to
be displayed e.g.
Lead III
First press
ST1 or ST2
Lastly press on
Cancel to clear
the changes or
Save & Back to
keep them,
either option will
return to the
ECG settings
menu