6 Taking medical readings
Note
When a medical module is first started the word “initialising” may appear for a few seconds
in the results area associated to the medical parameter being started.
6.1 Electrocardiography (ECG)
The Tempus Pro allows the user to either monitor the patient’s ECG or take a diagnostic recording of their
ECG (this is an optional feature). This section will explain how to use both features. The device defaults to
ECG monitoring with either a 3-lead, a 4-lead or a 5-lead cable.
WARNING
The ECG device is not intended for use in a sterile environment. Do not use for direct
cardiac application.
WARNING
The ECG device is reusable, other than disposable single-use electrodes.
WARNING
Do not attempt to insert the ECG device (including patient cables) into an electrical
outlet.
WARNING
The ECG is for resting recordings and should not be used in stress testing
environments.
WARNING
Ensure electrodes are connected only to the patient.
WARNING
Conductive parts of electrodes and connectors, including neutral electrode, should
not contact other conductive parts including earth.
WARNING
During ECG recordings, ECG alarms are not active and will only be reactivated once
monitoring is resumed.
WARNING
The Tempus Pro is rated as being proof against the
effects of a d
efibrillator
discharge. Follow these warnings if using an AED or defibrillator with the Tempus
Pro:
Follow the instructions of the defibrillator or AED when using it with the Tempus Pro.
Do not touch the patient during defibrillation.
Do not touch the defibrillator’s paddle-electrode surface when discharging the
defibrillator.
Keep defibrillation electrodes well clear of other electrodes or metal parts in contact with
the patient.