4 Setting up
4.1 Before deployment
Prior to deployment of Tempus Pro there are settings that user organisations need to configure. Carefully
check all the relevant settings on each Tempus Pro before it is deployed. When deploying multiple units,
you may set up a single unit and export the configuration from one unit to the others (cloning the settings).
Three types of configuration are exporting and imported:
events (summary record of care);
new patient defaults;
settings.
Summary record of care events include:
summary record of care assessment, intervention, drug and fluid events;
summary record of care injury and burns body map tags;
summary record of care general screen events selection.
New patient defaults include:
patient alarm limits for all parameters;
default patient mode (adult, paediatric or neonate);
alarm silence and suspend times;
default NIBP settings (mode, auto-time and inflation pressures);
default ECG settings (HR/PR source, ST/QT etc.).;
printer default settings.
Settings include:
communications settings (e.g. IP address);
communications menu;
email settings;
ePCR settings;
patient record type, language, passwords.
Note
Configuration needs to be carried out by a biomedical engineer or suitably qualified person
following the instructions in the Tempus Pro Configuration Utility User Guide.
Note
The unit name is not cloned from one device to another.
Note
Licensed options and enabled features are not cloned from one device to another.