3.2 Key features of the unit
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 29
additional protection to prevent nuisance ingress of sand into the connectors in the event that a cable or
sensor is not in place.
3.2.5 Tempus Pro right side
The right side of the Tempus houses the data communication and power connections. These comprise:
Two USB 1.0 & 2.0 sockets – these are reserved ONLY for non-mains powered radios (or other
communications devices), non-mains powered USB peripherals (such as keyboards) or non-mains
powered medical accessories. Any accessory or peripheral attached to the Tempus MUST be
approved for use with the Tempus Pro by RDT.
Mains Power – only use the Tempus Pro approved mains power supply part number 01-2049 provided
by RDT.
Audio – this is only for use with the Wired Headset (part number 01-1019) supplied by RDT or for use
with the Tactical Headset Adaptor Cable (part number 01-2041) supplied by RDT (this cable adapts to
Peltor® or similar military-style headsets).
Tactical Switch – this switch allows the user to switch the Tempus’ visual and audible outputs to a
minimal setting and is described in section 4.3.2.
WARNING
The USB connection must only be connected to non
-
mains powered periphe
rals
approved by RDT and documented in this manual, or to interface accessories
provided by RDT. Any connections made to the USB port must be to medical or IT
peripherals or communications systems which comply with the applicable IEC safety
standard (i.e. IEC60601-1 or IEC60950). Any connection arrangements must be made
in a manner compliant with IEC60601-1. Users must ensure that they prevent the
patient from coming into contact with the peripheral and also must ensure the user
does not touch the peripheral while touching the patient. Alternatively, a suitable
isolation device can be used. RDT recommends the use of appropriate opto-isolation
devices such as the Ulinx™ Model UH401 or the Baaske USB 1.1 isolator. Further
details can be obtained from RDT upon request.
Left side of the Tempus – medical connectors
Capnometer
door
Contact
Temperature
doors
Invasive
Pressure door
Pulse Oximetry
(SpO2) patient
cable
ECG cable
NIBP hose