7.9 Testing alarms
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 145
Once set, Tactical Mode can be turned off by switching the physical switch back to its original position – see
“4.3.2 Tactical mode (optional)”.
WARNING
Do not leave a unit in tactical mode without careful observation for alarm conditions
on the screen.
7.9 Testing alarms
RDT recommends that visual and audible alarms are tested before use.
The alarm bar will light and the audible alarm will sound on switch on (unless the tactical switch is enabled).
If users do not perceive this test occurring, they should provoke a test to check that the visual and audible
alarms are working before they begin to use the device. A simple way to do this would be to insert a finger
into the pulse oximeter probe, wait for a reading to appear and then remove the finger. This will provoke a
technical alarm which will test the alarm bar (will light solid yellow) and the alarm tone.