Operator's Manual

7.5 Setting alarms
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 137
Alarm
Nominal Time to Delay (Excluding User
Configurable Delay)
QT Alarm 60 seconds
ST Alarm 60 seconds
(*) The user configurable delay is not applicable to this alarm
7.5.2 All alarms menu
CAUTION
The Reset to defaults button resets the following to their new patient default values: all
alarm limits, Alarm trigger delay (Alarm suspend time), Alarm silence time, NIBP mode,
NIBP adult inflation, NIBP paediatric inflation, NIBP neonate inflation, HR/PR source,
Pacemaker indication, Arrhythmia analysis, ST/QT analysis, Monitoring filter, 12-lead filter
and ECG power filter.
Menu option Menu screen
Alarm trigger delay
– set a delay time to help
prevent spurious alarms. Range is 0-8
seconds (4 seconds default, 0 seconds
minimum).
Alarm volume – select Low, Mid or High
volume.
Waveform snapshot on alarm – turn this
feature on or off.
Alarm silence timeset the time period in
which audible alarms will be silenced when
the Alarm Silence button is pressed. The
available range is from 30 seconds up to the
maximum value set in the Maintenance Menu.
All alarms on – activate all alarms (present
on all pages of this menu).
Reset to defaults - reset all alarm limits and
other parameters to their new patient default
values.
Heart rate limits – set alarm limits.
ST1 limits – set alarm limits.
ST2 limits – set alarm limits.
QT upper limit – set alarm limit.
ECG Alarms – turn ECG alarms off or on.