Operator's Manual
Table Of Contents
Chapter 7 Alarms
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 136
Alarm
Nominal Time to Delay (Excluding User
Configurable Delay)
SpHb high 2 seconds
SpMet low 2 seconds
SpMet high 2 seconds
SpCO low 3 seconds
SpCO high 4 seconds
PI low 2 seconds
PI high 2 seconds
PVI low 2 seconds
PVI high 2 seconds
SpOC low 2 seconds
SpOC high 2 seconds
ETCO
2
high 4 seconds
ETCO
2
low 4 seconds
Respiration rate (from Capnometer) high 4 seconds
Respiration rate (from Capnometer) low 4 seconds
Respiration rate (from ECG) high 4 seconds
Respiration rate (from ECG) low 4 seconds
Capnometer no breaths detected
10
-
60 seconds
(depending on delay setting)
Non-invasive blood pressure systolic high 0 seconds at end of measurement
Non-invasive blood pressure systolic low 0 seconds at end of measurement
Non-invasive blood pressure mean high 0 seconds at end of measurement
Non-invasive blood pressure mean low 0 seconds at end of measurement
Non-invasive blood pressure diastolic high 0 seconds at end of measurement
Non-invasive blood pressure diastolic low 0 seconds at end of measurement
Contact temperature low 4 seconds
Contact temperature high
4
seconds
Invasive pressure systolic high 4 seconds
Invasive pressure systolic low 4 seconds
Invasive pressure diastolic high 4 seconds
Invasive pressure diastolic low 4 seconds
Invasive pressure mean low 4 seconds
Invasive pressure mean high 4 seconds
ECG monitoring arrhythmia - Extreme Brady
6 seconds (*)
ECG monitoring arrhythmia - Extreme Tachy
7 seconds
ECG monitoring arrhythmia - Vent Tachy
5 seconds
ECG monitoring arrhythmia - Vent Fib
9 seconds
ECG monitoring arrhythmia - Asystole
6 seconds