Operator's Manual
Table Of Contents
Chapter 7 Alarms
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 134
Condition On-Screen Error Text Comment
Bluetooth pairing error Bluetooth Pairing Error
Entered PIN is incorrect
Repeat the process taking care
to follow on-screen and written
instructions
Bluetooth pairing error Warning - Bluetooth
Bluetooth is not activated on this device.
Repeat the process taking care
to follow on-screen and written
instructions
System error Internal Fault Detected
A fault has been detected. Reason:
exception. The Tempus will need to restart.
If this problem persists please contact your
supplier.
Log the exact details of the fault
so they can be communicated to
the supplier.
System error Internal Fault Detected
A fault has been detected. Reason: file
open error. The Tempus will need to restart.
If this problem persists please contact your
supplier.
Log the exact details of the fault
so they can be communicated to
the supplier.
System error Internal Fault Detected
A fault has been detected. Reason: file
read error. The Tempus will need to restart.
If this problem persists please contact your
supplier.
Log the exact details of the fault
so they can be communicated to
the supplier.
System error Internal Fault Detected
A fault has been detected. Reason: file
write error. The Tempus will need to restart.
If this problem persists please contact your
supplier.
Log the exact details of the fault
so they can be communicated to
the supplier.
System error Internal Fault Detected
A fault has been detected. Reason:cOS
exception. The Tempus will need to restart.
If this problem persists please contact your
supplier.
Log the exact details of the fault
so they can be communicated to
the supplier.
IP address error Invalid IP Address
Please check the IP Address is in the
format x.x.x.x and multiple addresses
(maximum of 3) are separated by ';'.
Correct the format of the IP
address
Network Settings Error Network Settings Error
There are no network settings in this mode.
Enter network settings
Software update error Options Key Check Failed
One or more option key checks have failed.
Some features may be disabled. If the
problem persists contact the service
representative.
A software upgrade may have
occurred incorrectly. Log the
exact details of the fault so they
can be communicated to the
supplier.
Low ambient temperature
warning
Temperature Warning
The device temperature is low, Tempus
may shutdown.
Attempt to warm the device and
then restart