Operator's Manual
Table Of Contents
7.3 Patient alarms
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 131
Parameter Measurement
Range
Patient
type
Default
Alarm
Low
Alarm Low
Range
Default
Alarm
High
Alarm High
Range
Invasive
Pressure –
Systolic
Arterial
-99 – 310
mmHg
Adult 90 mmHg -99 – 310
mmHg
160 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 70 mmHg -99 – 310
mmHg
120 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 40 mmHg -99 – 310
mmHg
90 mmHg -99 – 310
mmHg
Invasive
Pressure –
Diastolic
Arterial
-99 – 310
mmHg
Adult 50 mmHg -99 – 310
mmHg
90 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 40 mmHg -99 – 310
mmHg
70 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 20 mmHg -99 – 310
mmHg
60 mmHg -99 – 310
mmHg
Invasive
Pressure –
Mean
Arterial
-99 – 310
mmHg
Adult 60 mmHg -99 – 310
mmHg
110 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 50 mmHg -99 – 310
mmHg
90 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 24 mmHg -99 – 310
mmHg
70 mmHg -99 – 310
mmHg
Invasive
Pressure –
Systolic
Venous
-99 – 310
mmHg
Adult 6 mmHg -99 – 310
mmHg
14 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 2 mmHg -99 – 310
mmHg
10 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 2 mmHg -99 – 310
mmHg
10 mmHg -99 – 310
mmHg
Invasive
Pressure –
Diastolic
Venous
-99 – 310
mmHg
Adult -4 mmHg -99 – 310
mmHg
6 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric -4 mmHg -99 – 310
mmHg
2 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate -4 mmHg -99 – 310
mmHg
2 mmHg -99 – 310
mmHg
Invasive
Pressure –
Mean
Venous
-99 – 310
mmHg
Adult 0 mmHg -99 – 310
mmHg
10 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 0 mmHg -99 – 310
mmHg
4 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 0 mmHg -99 – 310
mmHg
4 mmHg -99 – 310
mmHg
Invasive
Pressure –
Systolic
PAP
-99 – 310
mmHg
Adult 10 mmHg -99 – 310
mmHg
34 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Paediatric 24 mmHg -99 – 310
mmHg
60 mmHg -99 – 310
mmHg
-99 – 310
mmHg
Neonate 24 mmHg -99 – 310
mmHg
60 mmHg -99 – 310
mmHg
Invasive
Pressure –
-99 – 310
mmHg
Adult 0 mmHg -99 – 310
mmHg
16 mmHg -99 – 310
mmHg