Operator's Manual

Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 13
2 Warnings and cautions
2.1 Safety and note icons
This manual uses the following icons are used to indicate safety, caution and warning messages
(per ISO 3864-2):
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or
moderate injury.
The note icon is used for important and helpful information:
Note
A point of particular interest or emphasis intended to provide more effective or convenient
operation.
2.2 EMC information
The Tempus Pro Patient Monitor has been tested and approved to IEC/EN60601-1-2:2014. This means that
the Tempus meets or exceeds the requirements for electrical medical equipment in terms of its levels of
emitted electromagnetic (EM) radiation and its susceptibility to electromagnetic radiation from other devices.
In addition, the Tempus Pro has been tested according to the requirements of RTCA DO-160-G section 21
category M.
It should be noted that the Tempus Pro may be affected by high levels of stray EM radiation from other
electronic devices (even those which comply with relevant CISPR emission standards) that are being used
in close proximity to it.
As required by international medical device standards, the Tempus Pro is intended for use in
electromagnetic environments of 8kV static contact (15kV air discharge) and magnetic fields of 30 A/m
(50/60Hz). The Tempus Pro is proof against radiated RF emissions from 80MHz to 2.7GHz to a level of at
least 3V/m. In the event that the Tempus Pro will be used in environments with RF levels exceeding this,
please contact RDT for further information.