The table below lists all patient alarms:
Parameter Measurement
Range
Patient
type
Default
Alarm
Low
Alarm Low
Range
Default
Alarm
High
Alarm High
Range
ECG Heart
rate
30-300 bpm Adult 50 bpm 30-238 bpm 120 bpm 32-239 bpm
30-300 bpm Paediatric
50 bpm 30-238 bpm 150 bpm 32-239 bpm
30-300 bpm Neonate 100 bpm 30-238 bpm 200 bpm 32-239 bpm
ST -50 - +50 mm Adult -2.0 mm -10 – +2 mm +2.0 mm -2 – +10.0
mm
N/A Paediatric N/A N/A N/A N/A
N/A Neonate N/A N/A N/A N/A
QT 1 – 2000 ms Adult N/A N/A 500 ms 10 – 1990
ms
N/A Paediatric
N/A N/A N/A N/A
N/A Neonate N/A N/A N/A N/A
PI 0-20% Adult OFF Off, 0.1 to 18 OFF 0.2 to 19, Off
0-20% Paediatric
OFF Off, 0.1 to 18 OFF 0.2 to 19, Off
0-20% Neonate OFF Off, 0.1 to 18 OFF 0.2 to 19, Off
PVI 0-100% Adult 5% Off, 1 to 98 40% 2 to 99, Off
0
-
100%
Paediatric
5%
Off, 1 to 98
40%
2 to 99, Off
0-100% Neonate 5% Off, 1 to 98 40% 2 to 99, Off
SpOC 0 to 35 ml/dl Adult 10 ml/dl Off, 1 to 33 25 ml/dl 2 to 34, Off
0 to 35 ml/dl Paediatric 10 ml/dl Off, 1 to 33 25 ml/dl 2 to 34, Off
0 to 35 ml/dl Neonate 10 ml/dl Off, 1 to 33 25 ml/dl 2 to 34, Off
SpO
2
pulse
rate
25-239 bpm Adult 50 bpm 30-238 bpm 120 bpm 32-239 bpm
25-239 bpm Paediatric 50 bpm 30-238 bpm 150 bpm 32-239 bpm
25-239 bpm Neonate 100 bpm 30-238 bpm 200 bpm 32-239 bpm
Impedance
respiration
3-150 rpm Adult 5 rpm 3-145 rpm 30 rpm 5-149 rpm
3-150 rpm Paediatric 5 rpm 3-145 rpm 30 rpm 5-149 rpm
3-150 rpm Neonate 12 rpm 3-145 rpm 80 rpm 5-149 rpm
SpO
2
0-100% Adult 90% 50-98% 100% 52-100%
0-100% Paediatric 90% 50-98% 100% 52-100%
0-100% Neonate 85% 50-98% 95% 52-100%
SpCO 0-99% Adult OFF 1-97% 10% 2-98%
0-99% Paediatric OFF 1-97% 10% 2-98%
0-99% Neonate OFF 1-97% 10% 2-98%
SpHb 0-25 g/dl Adult 7.0 g/dl 1.0-23.5 g/dl 17.0 g/dl 2.0-24.5 g/dl
0-25 g/dl Paediatric 7.0 g/dl 1.0-23.5 g/dl 17.0 g/dl 2.0-24.5 g/dl