Operator's Manual

Chapter 6 Taking medical readings
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 118
CAUTION
Remember to zero the channel/transducer that you are using. Take care to differentiate
between the two transducers and their settings.
If the process fails to measure a valid zero reading, check the transducer stop cock is open to atmospheric
air and correctly connected to the Tempus. Then repeat the process. Replace the transducer if the error
recurs.
1. Ensure you
are zeroing the
correct channel
2. Press here to
zero the
transducer
3. A zeroing
message will
appear here –
this lasts a few
seconds
4. Once the zero
has been
achieved, the
text will update
to confirm the
time the zero
was performed
Zeroing the transducer