6.6 Invasive pressure
WARNING
DO NOT use the Invasive Pressure monitor for any purpose other than specified in
this manual.
WARNING
If electrocautery is used, always avoid using any transducer with a conductive (metal)
case connected to its ground shield. Using a conductive transducer that is
connected to its cable shield risks high-frequency burns at the ECG electrodes if the
transducer case becomes grounded to earth.
WARNING
Normal alarm functions will detect complete disconnections of invasive pressure
transducers and faulty cables / transducers. Use only approved transducers and
ensure that they are connected properly.
WARNING
Invasive Pressure is a tool to be used in addition to patient assessment. Care should
be taken to assess the patient at all times.
WARNING
Do not touch metal parts of the transducer while it is in contact with the patient.
WARNING
Do not reuse any components that are labelled for single use only.
WARNING
Although complete disconnections of invasive pressure transducers will be detected
by the normal alarm functions, partial disconnection will not be detected, nor will the
use of some incompatible transducers. The user must exercise reasonable measures
to ensure that approved transducers are used and that pressure transducers are
connected properly.
WARNING
If electrocautery is used, always avoid using any transducer with a conductive (metal)
case that is electrically connected to its cable shield. Using a conductive transducer
case with such a shield connection risks high-frequency burns at the ECG electrodes
if the transducer case becomes earth grounded.
WARNING
Always read and follow the instructions on the packaging of transducers regarding
their storage, use and disposal. In particular take into account any information
regarding toxicity.
WARNING
Remember to follow instructions provided with the transducers relating to use in
defibrillation.
WARNING
Remember to differentiate the different transducers that are connected to the
Tempus. The adaptor cables are labelled P1 & P2. The Tempus provides features to
additionally label these channels with their site description on the display.
CAUTION
Do not attempt to clean, disinfect, sterilize or in any way re-use transducers, catheters or
other patient connected tubing which are labelled as single use only.
CAUTION
Dispose of all single use items according to standard operating procedures or local
regulations for the disposal of contaminated medical waste.