Chapter 6 Taking medical readings
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 110
6.5.3 Contact temperature settings
Pressing anywhere on the contact temperature area brings up the contact temperature settings menu.
Note
The thermometer performs a self-check to 38.8 C ±0.1 C constantly to ensure that its
readings are accurate and within specification.
If the Tempus Pro is configured to support only one temperature channel:
The values in the Temperature Settings screen apply regardless of whether the temperature sensor is
connected to socket T1 or T2.
The maintenance user can still configure default alarm limits for two channels (T1 and T2), but the T2
default alarm limits are ignored and the initial temperature limits in the Temperature Settings screen are
equal to the T1 default alarm limits.
The single channel Temperature Settings screen looks like this:
Menu option Menu screen
Temperature units – set the measurement
units to C or F.
Temp limits – set the lower and upper
temperature alarm limits.
If the Tempus Pro is configured to support two channels of contact temperature:
There are separate configured alarm limits for the T1 and T2 channels.
There is a configured alarm limit for the difference between T1 and T2 (Delta).