6.5 Contact temperature
Tempus Pro User/Operator Manual - 41-2001EN-00 – Page 109
6.5 Contact temperature
WARNING
The thermometer is only intended for use with YSI series 400 sensors. Disposable
and reusable 400 series sensors may be used. Do not use series 700 sensors with
the Tempus.
WARNING
Application and use of metal
-
jacketed temperature sensors that come into contact
with conductive objects or clinical personnel during electrocautery may cause burns
at the patient-probe/electrode contact points.
WARNING
Always read and
follow the instructions on the packaging of temperature sensors
regarding their storage, use and disposal. In particular take into account any
information regarding toxicity.
Note
The contact temperature probes must be attached to the parts of the body which their
labelling shows they are indicated for.
Note
A single temperature channel (T1) is provided on the Tempus as standard. The second
channel may be activated as an option using a software update.
6.5.1 Getting started
To use the thermometer, plug a YSI series 400 temperature sensor into the Tempus’ ¼” socket marked “T1”.
6.5.2 Taking readings
Attach the sensor to the patient following the instructions on the sensor’s packaging.
The device will start to provide readings as soon as the sensor is connected (typically <10 seconds but up to
35 seconds according to IEC80601-2-56).
Users should monitor the temperature to observe when the reading settles (settling time will be dependent
on environmental and patient factors).
Change the setting of the Tempus to C or F as required.
Note
Ensure the probe is suitably secured to the patient’
The readings
from channel T1
are shown here
The reading from
channel T2 are
shown here
The difference
between
channels T1 and
T2 is shown here