User Manual
CELLTECH RESEARCH INC. Test Report S/N: 092601-169ARU
1955 Moss Court, Kelowna Date of Tests: October 16, 2001
B.C. Canada V1Y 9L3 FCC SAR Evaluation
_________________________________________________________________________________________________________
© 2001 Celltech Research Inc. RELM WIRELESS CORP. FCC ID: ARURPV516A 6
Portable VHF PTT Radio Transceiver (150-174MHz)
7.0 SYSTEM VALIDATION
Prior to the assessment, the system was verified in the planar section of the phantom with a 900MHz
dipole for devices operating below 1GHz, and an 1800MHz dipole for devices operating above 1GHz. A
forward power of 250mW was applied to the dipole and system was verified to a tolerance of +10%. The
applicable verifications are as follows (see Appendix B for validation test plot):
8.0 TISSUE PARAMETERS
The dielectric parameters of the fluids were verified prior to the SAR evaluation using an 85070C
Dielectric Probe Kit and an 8753E Network Analyzer. The dielectric parameters of the fluid are as
follows:
Dipole Validation Kit Target SAR 1g (w/kg)
Measured
SAR 1g (w/kg)
D900V2 2.78 2.76
TISSUE PARAMETERS FOR DIPOLE VALIDATION
Brain Equivalent Tissue
Dielectric Constant
εε
r
Conductivity
σσ (mho/m)
ρρ (Kg/m
3
)
900MHz (Target) 41.5 ±5% 0.97 ±5% 1000
900MHz (Measured)
10/16/01
42.4 0.97 1000
BRAIN TISSUE PARAMETERS - EUT EVALUATION
Equivalent Tissue
Dielectric Constant
εε
r
Conductivity
σσ (mho/m)
ρρ (Kg/m
3
)
150MHz Brain
(Target)
52.3 ±5% 0.76 ±5% 1000
150MHz Brain
(Measured: 10/16/01)
52.6 0.76 1000
BODY TISSUE PARAMETERS - EUT EVALUATION
Equivalent Tissue
Dielectric Constant
εε
r
Conductivity
σσ (mho/m)
ρρ (Kg/m
3
)
150MHz Body
(Target)
61.9 ±5% 0.80 ±5% 1000
150MHz Body
(Measured: 10/16/01)
62.2 0.80 1000