CELLTECH RESEARCH INC. Test Report S/N: 092601-169ARU
1955 Moss Court, Kelowna Date of Tests: October 16, 2001
B.C. Canada V1Y 9L3 FCC SAR Evaluation
_________________________________________________________________________________________________________
© 2001 Celltech Research Inc. RELM WIRELESS CORP. FCC ID: ARURPV516A 9
Portable VHF PTT Radio Transceiver (150-174MHz)
ET3DV6 E-Field Probe
12.0 PROBE SPECIFICATION (ET3DV6)
Construction: Symmetrical design with triangular core
Built-in shielding against static charges
PEEK enclosure material (resistant to organic solvents, e.g. glycol)
Calibration: In air from 10 MHz to 2.5 GHz
In brain simulating tissue at frequencies of 900 MHz
and 1.8 GHz (accuracy ± 8%)
Frequency: 10 MHz to > 6 GHz; Linearity: ± 0.2 dB
(30 MHz to 3 GHz)
Directivity: ± 0.2 dB in brain tissue (rotation around probe axis)
± 0.4 dB in brain tissue (rotation normal to probe axis)
Dynam. Rnge: 5 µW/g to > 100 mW/g; Linearity: ± 0.2 dB
Srfce. Detect. ± 0.2 mm repeatability in air and clear liquids over
diffuse reflecting surfaces
Dimensions: Overall length: 330 mm
Tip length: 16 mm
Body diameter: 12 mm
Tip diameter: 6.8 mm
Distance from probe tip to dipole centers: 2.7 mm
Application: General dosimetry up to 3 GHz
Compliance tests of mobile phone
13.0 SAM PHANTOM V4.0C
The SAM phantom V4.0C is a fiberglass shell phantom with a 2.0mm
shell thickness for left and right head and flat planar area integrated in a
wooden table. The shape of the fiberglass shell corresponds to the
phantom defined by SCC34-SC2. The device holder positions are
adjusted to the standard measurement positions in the three sections.
14.0 DEVICE HOLDER
The DASY3 device holder has two scales for device rotation (with respect
to the body axis) and the device inclination (with respect to the line
between the ear openings). The plane between the ear openings and the
mouth tip has a rotation angle of 65°. The bottom plate contains three pair
of bolts for locking the device holder. The device holder positions are
adjusted to the standard measurement positions in the three sections.
SAM Phantom
Device Holder