Operator`s manual

Q-DL-2100 Operatorʼs Manual
Quaesta Instruments, LLC (520) 449-3806 Email: support@quaestainstruments.com
16
2.4 Sample QILogger.ini file, cont
//Digital I/O Configuration
Dio1=1
Dio2=1
Dio3=3
Dio4=2
Digital I/O channels 1 and 2 are configured as counters while channels 3 and 4 are
configured as a Digital Output and a Digital Input, respectively.
// Neutron Pulse Module Configuration
NumberOfNPMs=2
The number of Neutron Pulse Modules present in the Probe is set to 2. The current
version of the datalogger firmware supports from 0 to 4 Neutron Pulse Modules.
// DAC Output Mapping
DACMappingOn=1
DAC1=P1* 0.003000
DAC2=T1* 0.08000
DAC3=N1* 0.01000
DAC4=N2* 0.01000
The DACMappingOn INI parameter allows one to produce Analog Outputs
proportional to Inputs sampled by the datalogger. i.e., with the parameters as above, Analog
Output Channel 1 will be a voltage output equal to (P1 in mbar)*0.003, Analog Output
Channel 2 will be a voltage output equal to (T1 in deg C)*0.08, Analog Output Channels 3
and 4 will be a voltage equal to Neutron Pulse Module Counts (N1*0.01) and (N2*0.01),
respectively.
*Note that the maximum value of the Analog Output voltage is 5V DC, so the scaling
constants used should be such that 5V is not obtained over the expected range of the
parameter to be mapped.
// Relays and Regulated Power Outputs
Relays=0
Pwr5VOn=0
Pwr12VOn=1
PwrNPMOn=1
All relays are OFF (deactivated).
5V Regulated Power Output is OFF (disabled).
12V Regulated Power Output is ON (enabled).
Neutron Pulse Module Power is ON (enabled).
// Power Conservation Configuration
// -----These params are active only if Battery Thresholding
// -----is On or Daily Power Conservation is On
Conserve5VPwr=1
Conserve12VPwr=1
ConserveNPMPwr=1
5V and 12V Regulated Power, as well as NPM Power, will be conserved if Battery
Thresholding or Daily Power Conservation is On