Operator`s manual

Q-DL-2100 Operatorʼs Manual
Quaesta Instruments, LLC (520) 449-3806 Email: support@quaestainstruments.com
12
2.3 INI Parameters Description
Q-DL-2100 INI file parameters (Not Case Sensitive)
Name= ? - Name of the Logger: e.g. 'name= QILogger'(16 max)
Extension= ? - Extension of the filename: e.g. 'extension= 005'(3 chars max)
*The extension of the data filenames written to the SD media
// Data Acq Params
StartTime= ? - Time to start logging (YYYY/MM/DD HH:MM:SS);
e.g. 'starttime= 2009/07/04 14:30:00
StopTime= ? - Time to stop logging (YYYY/MM/DD HH:MM:SS);
e.g. 'stoptime= 2019/01/01 01:00:00
RecordPeriod=? - Record period in minutes: e.g. 'Recordperiod=15'
*This is the interval at which data is sampled and recorded on the Secure
Digital card.
DataSelect= ? - Select Data to save:
e.g. 'dataselect= p1t1h1b1n1n2'
p1=Internal Press Sensor, t1=intern temp h1=intern RH,
b1=Power Supply Voltage,
n1=1st NPM cts, n2=2nd NPM cts, n3=3
rd
NPM cts, etc
t2=extern temp, h2=extern RH
aN=Analog Input Chan N, N= 1 to 4, e.g. a2 selects Analog Input Ch2
dN=Digital Input Chan N, N= 1 to 4, e.g. d3 selects Digital Input Ch3
//Analog Input Configuration Params
AdcCalN= ?.??? - N= ADC Chan #, Calibration Multiplier for ADC Channel N
e.g. 'AdcCal1= 1.3'
Pcal= ?.? - Barometric Pressure calibration offset, millibars: e.g. 'Pcal= 8.9'
* This is an additive number. i.e., if the Pressure is reading low by 10.2 mbars,
then set Pcal 10.2 mbars higher than its current value.
//Digital I/O Configuration
DioN= ? - N= Digital I/O Chan: ? -> 1=CTR mode, 2=Output, 3=Input.
e.g. ʻDio2=1ʼ
//Neutron Pulse Module Configuration
NumberOfNPMs= ? - Number of Neutron Pulse Modules (NPMs) in Use
// DAC Output Mapping
DACMappingOn= ? - Turns OFF/ON DAC Mapping
*makes DAC out proportional to given input)
DAC?=?? * ?.???? - Map Dac? value, e.g. 'Dac3= T1 * 0.128'.
*Active only if DACMappingOn=1.
// Relays and Regulated Power Outputs
Relays=? - Relays activated bitwise: e.g.
ʼRelays = 1ʼ, Relay 1 activated, Relay 2 deactivated
ʼRelays = 2ʼ, Relay 1 deactivated, Relay 2 activated
ʼRelays = 3ʼ, both Relay 1 and Relay 2 activated
ʼRelays = 15ʼ, all 4 relays activated
Pwr5VOn=? - Turns OFF/ON regulated 5V output
Pwr12VOn=? - Turns OFF/ON regulated 12V output
PwrNPMOn=? - Turns OFF/ON Neutron Pulse Modules