Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
Tono-Pen XL is suitable for use in electromagnetic environment specied below. The
customer or user of the
Tono-Pen XL should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete or ceramic
tile. If oors are synthetic, the R/H should be
at least 30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
N/A
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
N/A
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
N/A
Mains power quality should be that of a
typical residential, commercial or hospital
environment. If the user of the
Tono-Pen
XL
requires continued operation during power
mains interruptions, it is recommended that the
Tono-Pen XL be powered from an uninter-
ruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3 A/m
Power frequency magnetic elds should be that
of a typical residential, commercial or hospital
environment.
Guidance & Manufacturer’s Declaration (continued)