Instruction manual

195
Guidance and manufacturer’s declaration – electromagnetic
immunity - Part II
e T
Tono-Pen® XL
is intended for use in the electromagnetic environment specifi ed
below.  e customer or the user of the Tono-Pen should assure that it is used in such an
environment.
Immunity test
EN 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
N/A
3 V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the Tono-Pen,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in meters
(m).Field strengths from fi xed RF transmitters, as determined by an electromagnetic site
survey
a
, should be less than the compliance level in each frequency range
b
(m).Field strengths from fi xed RF transmitters, as determined by an electromagnetic site
b
(m).Field strengths from fi xed RF transmitters, as determined by an electromagnetic site
.
Interference may occur in the vicinity of equipment marked with the following symbol:
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2  ese guidelines may not apply in all situations. Electromagnetic propagation is
aff ected by absorption and refl ection from structures, objects and people
a
Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fi xed RF transmitters, an electromagnetic site survey should be
considered. If the measured fi eld strength in the location in which the Tono-Pen is used
exceeds the applicable RF compliance level above, the
Tono-Pen® XL
should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the
Tono-Pen® XL
.
b
necessary, such as re-orienting or relocating the
b
necessary, such as re-orienting or relocating the
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.