Instruction manual
27
SECTION 5
Maintenance
5.1 Maintenance Procedures
Maintenance that should be performed on the Tono-Pen XL
con
sists of activities such as protecting the probe tip, keep-
ing surfaces free of dust and dirt and storing in a dry and
cool place so as not to adversely aff ect electronic parts, and
replacing the batteries.
1. Keep the probe tip protected with an Ocu-Film Tip
Cover when in use or in storage. Avoid any contact with
the probe tip when not protected by an Ocu-Film Tip
Cover.
CAUTION
Use or storage of the Tono-Pen XL unit without an Ocu-Film
Tip Cover can cause severe damage to the instrument and
void
the Warranty.
2. e probe post must be cleaned only with optical qual-
ity
compressed gas before the fi rst use each day, before
instrument storage, in the event of unanticipated readings
or if a “Good” calibration check cannot be obtained. To
clean the probe post:
a. Remove the Ocu-Film Tip Cover.
b. Spray the compressed gas onto the probe tip for approxi-
mately 2 seconds.
c. Wait 3 minutes to allow the instrument to thermally
stabilize.
d. Cover the Tono-Pen XL unit’s probe tip with a new
Ocu-Film Tip Cover.