Instruction manual

25
e display of one of four horizontal bars located along the
lower border of the LCD block indicates the statistical reli-
ability of the averaged measurement. For example, if the 5%
bar is displayed, the standard deviation of the valid measure-
ments is 5% or less of the number shown.
If the reliability measure is 20% or > 20%, a repeat measure-
ment is recommended.
A single row of dashes [- - - -] indicates that the instrument
is activated.
Two rows of dashes [====] followed by a “beepindicates
that the instrument is ready to take measurements.
CAL
indicates that the instrument is in the process of cali-
bration check or the instrument needs to undergo calibration
check procedure. If “CALis displayed following a long beep
and “- - - -”, the latter is the case. A “Good” calibration check
must be obtained.
Lob
indicates the need to replace both batteries prior to
using the instrument (see Section 3).
UP
indicates that the instrument is being calibrated and
the probe tip should be turned from pointing down to point-
ing up.
Good
indicates the calibration check procedure was suc-
cessful. A er two Good” calibration checks, the Tono-Pen
XL unit is ready for use in patient examinations.
bAd
” indicates the calibration check procedure was unsuc-
cessful and must be repeated.