Instruction manual

23
occurs, repeat the patient examination procedure, starting at step 1.
9. To take another measurement, reactivate the Tono-Pen
XL unit by pressing and releasing the activation switch
as described in step 5.
*Note: If not reactivated within 25 seconds, the Tono-Pen XL
unit will automatically turn off and the LCD will clear, plac-
ing the device in sleep mode, to conserve battery life. If this
occurs, activate the Tono Pen XL by pressing the Operators
Button once.
10. Replace the Ocu-Film Tip Cover before using the Tono-
Pen XL unit on another patient and before storage.
*Note: Failure to replace the Ocu-Film Tip Cover
between patients may result in unreliable readings of IOP by
allowing debris to enter the transducer assembly.
*Note: A reused Ocu-Film Tip Cover may leak.  is may
cause damage to the strain gauge and void the Warranty.