Instruction manual

18
CAUTION
Do not use the instrument without an Ocu-Film Tip Cover.
Use without an Ocu-Film Tip Cover will damage the instru-
ment and void the Warranty.
4.2 Calibration
e Tono Pen XL unit is internally calibrated, thus the instru-
ment calibration should be checked only before the fi rst use
each day or in the event of unanticipated readings. Calibra-
tion must be routinely and successfully performed once daily
prior to instrument use. Calibration should also be performed
whenever it is indicated by the LCD display, when batteries
are replaced, or a er an unsuccessful calibration. It is not
necessary to check calibration prior to each use.
*Note: If during operation the Operator’s Button is pushed only
once and the Tono-Pen displays CAL, the Tono-Pen needs to
be calibrated.
e calibration procedure is as follows:
Point the transducer end of the Tono-Pen straight
down
towards the fl oor.
Depress the Operators Button two times rapidly, (within
1.5 seconds). e Tono-Pen will beepand display
CAL
.
Wait approximately 15 seconds for the Tono-Pen to
“beep” and display
UP
.
Immediately (within 1 second) invert the Tono-Pen
smoothly, pointing the transducer end straight
up
up
.
A properly functioning Tono-Pen will display Good
followed by a “beep.
Repeat this Calibration Procedure if
bAd
is displayed.
1.
2.
3.
4.
5.
6.