3.3 Calibration Procedure
Calibration must be performed whenever batteries are re-
placed or a er an unsuccessful calibration. e calibration
procedure must also be routinely and successfully performed
once daily prior to instrument use, or whenever indicated by
the LCD display. To calibrate the Tono-Pen XL:
Point the transducer end of the Tono-Pen straight
down
towards the fl oor (see Figure 3-6).
Depress the Operator’s Button two times rapidly, (within
1.5 seconds). e Tono-Pen will “beep” and display
CAL.
Wait approximately 15 seconds for the Tono-Pen to
“beep” and display
UP
(see Figure 3-6).
Immediately (within 1 second) invert the Tono-Pen
smoothly, pointing the transducer end straight
up
up
.
A properly functioning Tono-Pen will display Good
followed by a “beep”.
A properly functioning Tono-Pen will display Good
followed by a “beep”.
A properly functioning Tono-Pen will display Good
Repeat this Calibration Procedure if bAd is displayed.
A er Good is displayed, depress the Operator’s Button
one time and the Tono-Pen will display [8.8.8.8], followed
by a single row of dashes, [- - - -], and then by a double
row of dashes [====] followed by a “beep” tone, indicat-
ing the instrument is ready to measure IOP (Depressing
the activation button a second time will interrupt this
process and send the unit to calibration mode).
*Note:
If the Tono-Pen XL fails to yield a “Good” calibration,
call Medtronic Technical Service.