Instruction manual

11
SECTION 3
Set Up
3.1 Set Up Instructions
To set up the Tono-Pen XL for operation, follow the instruc-
tions listed below.
Remove all protective packaging materials. Remove and
discard the desiccant.
Carefully remove the tonometer and accessories from
the case.
Check for missing items.
Visually check the instrument and accessories for dam-
age.
Remove the Ocu-Film Tip Cover shipped with the
Tono-Pen.
Notify Medtronic Customer Service (see Section 7,
Warranty and Repairs) immediately if any components
are missing or damaged.
Install the batteries in the battery compartment.
Perform the Calibration Procedure (see Section 3.3).
Slide a new Ocu-Film Tip Cover over the Tono-Pen XL
units stainless steel probe until the Ocu-Film ridge is
seated in the groove (see Figure 3-1).  e Ocu-Film
Tip Cover should t over the probe tip (see Figure 3-2),
without being too tight. Proper t makes contact with
the transducer surface (probe post), but still has pleats
along the side of the transducer head assembly.
1.
2.
3.
4.
5.
6.
7.
8.
9.