Guidance & Manufacturer’s Declaration
Table 201 – Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The
Tono-Pen VET is intended for use in the electromagnetic environment specied below.
The customer or user of the
Tono-Pen VET should ensure that it is used in such an environ-
ment.
Emissions Test Compliance
Electromagnetic Environment
- Guidance -
RF Emissions
CISPR 11
Group 1
Class B
The
Tono-Pen VET uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
Harmonics
IEC 61000-3-2
N/A
The
Tono-Pen VET is suitable for use in all estab-
lishments, including domestic establishments and those
directly connected to the public low-voltage power sup
-
ply network that supplies building for domestic power.
Flicker
IEC 61000-3-3
N/A
Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
Tono-Pen VET is suitable for use in electromagnetic environment specied below. The cus-
tomer or user of the Tono-Pen VET should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete or ceramic
tile. If oors are synthetic, the R/H should be
at least 30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
N/A
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
N/A
Mains power quality should be that of a
typical residential, commercial or hospital
environment.
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
N/A
Mains power quality should be that of a typical
residential, commercial or hospital environ
-
ment. If the user of the Tono-Pen VET
requires continued operation during power
mains interruptions, it is recommended that the
Tono-Pen VET be powered from an uninter-
ruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3 A/m
Power frequency magnetic elds should be that
of a typical residential, commercial or hospital
environment.