User`s guide
24 16012-101 Rev. C
Cleaning
External Cleaning
WARNING: DO NOT IMMERSE THE connector OR THE main unit in ANY liquid OR DAMAGE TO THE UNIT WILL OCCUR.
Clean the external surfaces of this instrument using a clean, soft cloth moistened with a mild detergent solution (1 cc of
liquid dish soap to one liter of clean, filtered water (filtered below 5 microns)).
Probe Cleaning
WARNING: In order to prevent patient-to-patient transfer of infection, after each use disinfect the probe following accepted local
clinical procedures regarding the use of disinfectants. Any clinically approved chemical disinfectant can be used.
WARNING: Do not attempt to use the IOPac if there is any indication the Probe tip, body, or cable has been damaged and/or their
physical integrity has been compromised OR INJURY TO THE CORNEA MAY OCCUR.
CAUTION: DO NOT use the probe as a stylus OR DAMAGE TO THE PROBE MAY OCCUR.
CAUTION: DO NOT STERILIZE THE PROBE USING ETO, AUTOCLAVING, OR IRRADIATION OR DAMAGE TO THE PROBE
WILL OCCUR.
caution: If the probe tip has made contact with anything between applanations, repeat the cleaning procedure.
Clean the probe using the following steps:
1. Wipe the complete probe with a cotton swab soaked in 70% isopropyl alcohol.
2. Immerse the probe tip for 10 minutes in 70% isopropyl alcohol or an equivalent locally approved disinfectant.
3. Rinse the end of the probe thoroughly with sterile distilled water.
4. Dry the probe with a clean, lint-free, cloth or tissue.
5. Store the probe in its provided holder on the unit when it is not in use so that it is safely stored.