User`s guide

16012-101 Rev. C 19
System Menu Screen
The system menu screen is accessed by touching the “Reichert Pachymetry” menu at the top and left of the screen with
the stylus. When the “Reichert Pachymetry” menu is selected, the following options are shown. The details of each
menu selection are indicated below:
Check Calibration
Delete Database
IOP Formulas
IOP Lookup Table
IOP Reference List
Glaucoma Risk Calculator
About Pachymeter
Check Calibration
It is recommended that you check the calibration approximately every 14 days.
From the main screen, tap Reichert Pachymeter located on the upper left corner of the IOPac screen with the stylus.
Select: Check Calibration from the pull down menu.
Place a drop of water (Do not use ultrasonic gel) on the calibration phantom, tap the Calibrate button, and then place
the probe lightly on the calibration phantom. A tone signals calibration is complete and the results of the calibration are
reported.
Note: The probe must be perpendicular to the calibration phantom therefore some manual adjustment of probe
alignment may be required for the IOPac to obtain a reading. If the measured thickness is slightly greater or less
than the allowable range, try redoing the calibration while exerting less pressure onto the phantom. If the
measured thickness consistently measures outside of the tolerance range, please contact an authorized
Reichert dealer.