R7800 Coating Thickness Gauge Instruction Manual Find Quality Products Online at: www. GlobalTestSupply.com sales@GlobalTestSupply.
Table of Contents Introduction.................................................................................................... 3 Product Quality............................................................................................... 3 Safety............................................................................................................. 4 Features.......................................................................................................... 4 Included.......................
Zero Calibration.......................................................................................14-16 Zero Calibration....................................................................................... 14 Two-Point Calibration.............................................................................. 15 Calibration Verification............................................................................. 16 Locking Screen Auto-Rotation............................................................
Safety Never attempt to repair or modify your instrument. Dismantling your product, other than for the purpose of replacing batteries, may cause damage that will not be covered under the manufacturer's warranty. Servicing should only be provided by an authorized service center.
Specifications Measuring Range: Accuracy: Resolution: Measurement Modes: Display: Resolution: Backlit Display: Min/Max/Average Readings: Minimum Curvature Radius: Minimum Area Diameter: Minimum Substrate Thickness: Auto Shut-Off: Low Battery Indicator: Power Supply: Internal Memory: PC Connectivity: Software: Product Certifications: Refresh Rate: Operating Temperature: Storage Temperature: Operating Humidity Range: Dimensions: Weight: Find Quality Products Online at: 0 to 1250µm 0.1 to 49.
Instrument Description 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 1 2 Wrist Strap Anchoring Point LED Status Indicator LCD Display Power Button Menu/Confirm/Calibration Button Increase Value/Up/Lock Screen Button USB Port Cancel/Clear/Return Button Decrease Value/Down/Quick Decision Mode Button Battery Compartment (Back) Sensor R7800 3 7 Coating Thickness Gauge 4 8 ZERO OK 5 MENU 9 6 10 11 Display Description 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12.
Operating Instructions Power ON/OFF To turn the meter ON or OFF press and hold the power button for 2 seconds. Note: If the buzzer is enabled, there will be an audible beep when the instrument is powered ON or OFF. Measuring Procedure Before each test session perform a Zero Calibration. See the Zero Calibration section for details. 1. During start-up the meter begins an initialization process and enters normal operation upon completion. 2.
Setup Mode 1. Press the button to enter Setup Mode. OK ZERO MENU MENU um/mil 2. Use the and arrows to scroll through the following parameters.
Enabling/Disabling Audible Beep 1. 2. 3. 4. Press the OK button when the audible beep icon is selected in Setup Mode. and arrows to select between ON or OFF X. Use the Press the OK button to confirm selection. Press the button to exit Setup Mode and resume normal operation. OK OK ZERO MENU Setting the LCD Brightness 1. OK MENU OK ZERO MENU Press the OK button when the brightness icon is selected in Setup Mode. Use the and arrows to increase or decrease the brightness level.
Setting Higher Limit Value Press the OK button when the higher limit value icon is selected in Setup Mode. 2. Use the and arrows to increase or decrease the higher limit value. 3. Press the OK button to confirm selection. button to exit Setup Mode and resume normal 4. Press the operation. Note: If the higher limit value is exceeded while in normal measuring mode and LED indicators are enabled (See Enabling/Disabling LED Indicators for details), the LED warning light will flash yellow.
Enabling/Disabling LED Indicators Press the OK button when the LED alarm icon is selected in Setup Mode. Use the and arrows to turn the LED alarm ON or OFF. Press the OK button to confirm selection. Press the button to exit Setup Mode and resume normal operation. 1. 2. 3. 4. OK ZERO MENU OK ZERO MENU Factory Reset Press the OK button when the factory reset icon is selected in Setup Mode. Use the and arrows to turn the factory reset ON or OFF. Press the OK button to confirm selection.
Delete Recorded Data 1. 2. 3. 4. Press the OK button when the delete recorded data icon is selected in Setup Mode. Use the and arrows to turn the delete recorded data mode ON or OFF. Press the OK button to confirm selection. button to exit Setup Mode and resume normal operation. Press the OK ZERO MENU OK ZERO MENU Maximum, Minimum, and Average Values OK MENU OK ZERO MENU ZERO Minimum, maximum and average values are displayed during the measuring process.
Single-Point Decision Configuration 1. Press the and arrows to set the target thickness value, and press the button to confirm selection. Press the and arrows to set the tolerance value, and press the button to confirm. Use the gauge to measure the coating thickness of the measured object. The screen displays the measured value and whether or not the thickness passes or fails ("PASS" or "FAIL"). button to return to the single-point analyzing setup Press the button to exit and resume normal operation.
Zero Calibration While in normal operation press and hold the button to enter the selected calibration mode from the Setup Mode. See Setting Calibration Mode for details. OK MENU Calibration Mode Icon ZERO Description Zero Calibration Simply place the sensor on an uncoated metal substrate for zero calibration. Two-Point Calibration Place a standard coating thickness film OK with a known thickness on an uncoated ZERO substrate for calibration to obtain a more accurate measurement result for testing.
Two-Point Calibration 1. When Figure 3 appears on the screen, place a standard coating thickness film with known thickness (e.g.: 500μm) on top of an uncoated substrate. 2. Place the gauge vertically on both the thickness film and uncoated substrate. 3. Lift the instrument after approximately 2 seconds and the measured value will be displayed (as shown in Figure 4). 4. Press the and arrows to adjust the value to the actual thickness of the standard coating thickness film (as shown in Figure 5). 5.
Calibration Verification While in normal operation, measure the standard coating thickness film. The measured value of the instrument should be within the indicated accuracy on the coating thickness film. If it is out of tolerance, recalibration is required. Note: If the calibration result is inaccurate try restoring the factory settings and recalibrating. Locking Screen Auto-Rotation The instrument has a built-in gravity sensor, which automatically rotates the display during testing.
Battery Replacement When the " " icon appears on the LCD, the batteries should be replaced. 1. Remove the screw with a Philips screwdriver and open the battery cover. 2. Replace the 2 x "AA" batteries. 3. Secure the battery cover back and tighten the screw.
Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance. REED Instruments total liability is limited to repair or replacement of the product.
Test & Measure with Confidence Check out our latest Products! Find Quality Products Online at: www. GlobalTestSupply.com sales@GlobalTestSupply.