Full Product Manual
6
7. To take a measurement, position the sensor so that it is directly
touching the surface being tested. The moisture sensor ball head
must be kept perpendicular to the material being measured, pressed
rmly on the surface and not slanted.
8. TheLCDdisplaysarelativemoisturereadingfrom0.0%to100.0%
withprogrammablealarmindicators(DRY,RISKandWET)–See
"AlarmSetupMode"toprogramalarms.
9. The MAX and MIN indicators display the highest and lowest readings
recorded during the measuring period.
10. To hold the current measurement press the MEAS button and the
dataholdsymbolwillappearontheLCDscreen.Inordertoresume
normal operation, press the MEAS button again.
11. After60secondsofinactivity(inDataHoldMode),themeterwill
automatically power off.
Alarm Settings
1. ThedefaultRISKalarmvalueissetat30%.TheRISKalarmvalue
canbesetbetween0%and50%.–See"AlarmSetupMode"to
program alarms.
2. ThedefaultWETalarmvalueissetat60%.TheWETalarmvalue
canbesetbetween50%and100%.–See"AlarmSetupMode"to
program alarms.
Alarm Setup Mode
1. ToadjustRISKandWETalarmvalues,themetermustbeturnedon
andcalibrated(seesteps1to4inOperatingInstructions).
2. Press the MEAS button to put the meter in data hold mode and
press the SET button.
3. TheRISKsymbolwillappearontheLCDscreenaswellasthevalue
inwhichtheRISKalarmissetto.
4. To change the alarm value, press the UP or DOWN buttons.
5. Press the SET button to save your setting and to move onto the WET
alarm value.
6. TheWETsymbolwillappearontheLCDscreenaswellasthevalue
in which the WET alarm is set to.
continued...
REED Instruments
1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.com