19
3. Press the SETUP button and the "Range" symbol will ash on
the display.
4. Adjust the range setting by pressing the or buttons
5. Press the ENTER button save the range setting and proceed to the
value setting.
6. Press the or buttons to select the ashing digit required for
adjustment. Press the or button to adjust the value of the digit.
7. Press the ENTER button to save the value setting and proceed to the
Tolerance setting.
8.
Press the or buttons to scroll through following tolerance % values.
1)
±0.25%
2) ± 0.5%
3) ± 1%
4) ± 2%
9. Press the ENTER button to save the Tolerance setting.
10.
The upper display will now indicate "PASS" or "FAIL", depending on
whether the impedance measured exceeds tolerance range. The current
measurement result will display on the lower display.
11. Press the SORTING Button to exit Sorting Mode.
Battery Replacement
The battery icon indicates the power status of the batteries. As the
batteries weaken, the number of lines will decrease. When the battery icon
appears on the LCD, the batteries must be replaced.
To replace the batteries:
1. Remove the four screws under the ip stand located at the rear of
the meter.
2. Lift up the battery cover and replace the 6 x "AA" batteries
3. Secure the cover back into place with the four screws.
5) ±5%
6) ±10%
7) ±20%
8) -20% +80%