4. Press the CAL button again and a 30 second countdown will appear
on the screen.
5. After the countdown is complete, a "PASS" or "FAIL" will show on
the display.
6. Press the CAL button a third time, and "SRT" will appear on the LCD
screen, indicating to the user to put the shorting bar across the "+"
and "-" input terminals.
7. Press the CAL button a fourth time, and a 30 second countdown will
appear on the screen.
8.
After the calibration procedure is nished a "PASS" or "FAIL" will appear
on the LCD screen.
9.
Press the CAL button a fth time and the calibration data will be saved
if "PASS" was displayed for both the Open & Short calibrations.
Open Calibration (left) and Short Calibration (right)
Sorting Mode
Sorting mode helps the user quickly sort through multiple components.
Select the primary measurement mode (L/C/R) based on the type of
components to be measured. Insert the component to be used as the
standard or good reference value that will be used for testing against all
other components.
Note: The meter needs a component connected to either the input sockets
or terminals for sorting mode to activate.
1. When ready, turn the meter on and press the SORTING button to
enter sorting mode.
2. The upper display will indicate "PASS" and the lower display will
indicate the value of the component.
Note: The default sorting % is +/- 1%.