R5001 LCR Meter Instruction Manual REED Instruments 1-877-849-2127 | info@reedinstruments.com | www.reedinstruments.
Table of Contents Introduction................................................................................................. 3 Product Quality............................................................................................ 3 Safety.......................................................................................................... 4 Features....................................................................................................... 4 Included...................................
Relative Mode....................................................................................... 16 Accuracy Discrepancies..................................................................16-17 Capacitance..................................................................................... 16 Inductance....................................................................................... 16 Resistance.......................................................................................
Safety This manual should be read and well understood before the instrument is used. • Never attempt to repair or modify your instrument. Dismantling your product, other than for the purpose of replacing batteries, may cause damage that will not be covered under the manufacturer's warranty. Servicing should only be provided by an authorized service center. • Do not operate in an explosive atmosphere. • Do not operate in the presence of flammable gases or fumes. • Keep away from live circuits.
Included • • • • Test Leads and Probes Tweezers Alligator Clips Batteries Specifications Test Frequencies: Test AC Signal Level: Test Range (F=1kHz): Primary Parameters Display: Second Parameter Display: Power Supply: Dimensions: Weight: 100/120/1k/10k/100kHz 0.
Accuracy Notes: • Measurement performed at the test socket. • Measurements performed after correct open and short calibration. • DUT and test leads must be properly shielded. • Q value is the reciprocal of DF. • Accuracies based within 10% to 100% of full scale of range; values outside of range should be used as reference only. • "---" means parallel or series measurement mode. Inductance @ Ta = 18 to 28°C (De) Frequency = 100 Hz/120 Hz Range Resolution Lx Accuracy 20.000mH 1µH 1.
Frequency = 10kHz Range Resolution Lx Accuracy DF Accuracy 200.00µH 0.01µH 1.8% ±10d 1.8% ±50d Measurement Mode Series 2000.0uH 0.1µH 1.5% ±10d 1.5% ±50d Series Series 20.000mH 1µH 1.2% ±10d 1.2% ±50d 200.00mH 0.01mH 1.5% ±15d 1.5% ±50d --- 2000.0mH 0.1mH 2.0% ±10d 2.0% ±50d Parallel 20.000H 1mH 2.5% ±15d 2.5% ±50d Parallel Frequency = 100kHz DF Accuracy Measurement Mode 20.000µH Range Resolution Lx Accuracy 0.001µH 2.5% ±10d 2.5% ±50d Series 200.00µH 0.01µH 1.
Frequency = 1kHz Range Resolution Cx Accuracy DF Accuracy Measurement Mode 2000.0pF 0.1pF 3.5% ±15d 3.5% ±50d Parallel 20.000nF 1pF 1.0% ±10d 1.0% ±50d --- 200.00nF 0.01nF 0.9% ±10d 0.9% ±50d --- 2000.0nF 0.1nF 1.0% ±10d 1.0% ±50d Series 20.000µF 1nF 1.2% ±15d 1.2% ±50d Series 200.00µF 0.01µF 2.5% ±10d 2.5% ±50d Series 2000µF 1µF 4% ±20d 4% ±50d Series Frequency = 10kHz Range Resolution Cx Accuracy DF Accuracy Measurement Mode 200.00pF 0.01pF 3.0% ±8d 3.
Resistance @ Ta = 18 to 28°C (De) Frequency = 100Hz/120Hz Range Resolution Rx Accuracy Measurement Mode 200.00Ω 0.01Ω 1.2% ±10d --- 2.0000kΩ 0.1Ω 0.8% ±5d --- 20.000kΩ 1Ω 0.9% ±5d --- 200.00kΩ 0.01kΩ 0.7% ±3d --- 2.0000MΩ 0.1kΩ 1.0% ±5d --- 20.000MΩ 1kΩ 2.2% ±10d --- 200.0MΩ 0.1MΩ 2.5% ±10d --- Frequency = 1kHz Range Resolution Rx Accuracy Measurement Mode 20.000Ω 1mΩ 1.2% ±10d --- 200.00Ω 0.01Ω 0.8% ±5d --- 2.0000kΩ 0.1Ω 0.8% ±3d --- 20.000kΩ 1Ω 0.
Frequency = 100kHz Range Resolution Rx Accuracy 20.000Ω 1mΩ 2.3% ±10d Measurement Mode --- 200.00Ω 0.01Ω 1.5% ±5d --- 2.0000kΩ 0.1Ω 0.8% ±20d --- 20.000kΩ 1Ω 0.8% ±20d --- 200.00kΩ 0.01kΩ 1.5% ±10d --- 2.000MΩ 1kΩ 2.5% ±30d --- DC Resistance @ Ta = 18 to 28°C (De) Frequency = 100Hz/120Hz/1kHz/10kHz/100KHz Range Resolution Rx Accuracy 200.00Ω 0.01Ω 1.8% ±10d Measurement Mode --- 2.0000kΩ 0.1Ω 0.6% ±20d --- 20.000kΩ 1Ω 0.6% ±10d --- 200.00kΩ 0.01kΩ 0.5% ±3d --- 2.
Instrument Description R5001 10 11 3 4 12 13 14 5 6 7 15 16 8 9 1. 2. 3. 4. 5. 6. LCD Display Function Button Calibration Button Sorting Button Frequency Button Secondary Display/ Left Button 7. Power Button 8. Relative/Down Button 9. Input Sockets (Banana Jack Inputs) and Terminals for Positive, Negative, and Guard 10. 11. 12. 13. 14. 15. 16.
Operating Instructions Power ON/OFF To turn the meter on or off, press the power button. By default, the meter is set to Auto LCR Smart mode with a test frequency of 1kHz. Auto Power Off To preserve battery life, the meter is programmed to turn off after 5 minutes of inactivity and 3 warning beeps. If a button is pushed during the three beeps, the meter will continue measuring. To turn this feature on and off, press the APO button. When enabled, "APO" will appear on the display.
Test Frequency Functions Test frequency can greatly affect the results of measurement readings, especially when measuring inductors and capacitors. This section provides some recommendations and suggestions to consider. Capacitance When measuring capacitance, selecting the right frequency is important in obtaining the most accurate results. Generally a 1kHz test frequency is used to measure capacitors that are 0.01μF or smaller. For capacitors that are 10μF or larger, a lower frequency of 120Hz is used.
Primary Impedance with Secondary Parameter Test The default test mode for this meter is Auto-LCR Mode, which checks the type of impedance and automatically enters L/C/R Mode. The secondary parameter will follow the L/C/R measurement, meaning that (L+Q), (C+D)*, (R+θ)** are combined in one group respectively. 1. Press the FUNC button to switch from Auto-LCR Mode to Auto-Lp Mode, Auto-Cp Mode, Auto-Rp Mode, and DCR Mode.
Capacitance Most capacitance measurements will obtain best results with parallel mode selected. The majority of capacitors have a very low dissipation factor (high internal resistance) compared to the impedance of the capacitance. In these cases, the paralleled internal resistance has negligible impact on the measurement. There will be some cases where series mode would be used for a capacitance measurement, or the readings will appear inaccurate.
Relative Mode 1. To reserve the current DUT readings (DCUR) on the primary display as a reference value (DREF), press the button. 2. The ∆ symbol will then appear on the LCD screen indicating this function is now active. The secondary display will show the percentage of the relative value (REL%). 3. Press the button again to show the reference value (DREF) on the primary display; the ∆ symbol will start to blink. REL% = (DCUR – DREF) / DREF x 100% Note: The percentage range is -99.9% to 99.9%.
Resistance When measuring resistance of devices, it is important to know that there are two types or ways of measurement. One type is DC resistance measurement. Another type is AC resistance measurement. The LCR meter provides both types of measurement. Before using the meter to measure resistance, please verify whether the DUT (device under test) requires DC or AC resistance measurement method. Depending on the method, results will vary greatly.
4. 5. 6. 7. 8. 9. Press the CAL button again and a 30 second countdown will appear on the screen. After the countdown is complete, a "PASS" or "FAIL" will show on the display. Press the CAL button a third time, and "SRT" will appear on the LCD screen, indicating to the user to put the shorting bar across the "+" and "-" input terminals. Press the CAL button a fourth time, and a 30 second countdown will appear on the screen.
Press the SETUP button and the "Range" symbol will flash on the display. 4. Adjust the range setting by pressing the or buttons 5. Press the ENTER button save the range setting and proceed to the value setting. 6. Press the or buttons to select the flashing digit required for or button to adjust the value of the digit. adjustment. Press the 7. Press the ENTER button to save the value setting and proceed to the Tolerance setting. 8. Press the or buttons to scroll through following tolerance % values.
Accessories and Replacement Parts • R1210 Alligator Clip Set • R1020 Fused Test Lead Set • R8888 Deluxe Hard Carrying Case • CA-05A Soft Carrying Case Don't see your part listed here? For a complete list of all accessories and replacement parts visit your product page on www.reedinstruments.com. Product Care To keep your instrument in good working order we recommend the following: • Store your product in a clean, dry place. • Change the battery as needed.
Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one (1) year from date of shipment. During the warranty period, REED Instruments will repair or replace, at no charge, products or parts of a product that proves to be defective because of improper material or workmanship, under normal use and maintenance. REED Instruments total liability is limited to repair or replacement of the product.
Product Support If you have any questions on your product, please contact your authorized REED distributor or REED Instruments Customer Service by phone at 1-877-849-2127 or by email at info@reedinstruments.com. Please visit www.REEDINSTRUMENTS.com for the most up-to-date manuals, datasheets, product guides and software. Product specifications subject to change without notice. All rights reserved.