Full Product Manual
Data Logging
Data Recording Modes
Manual Data Logging: Press the LOG button to manually log up to
99 readings on a SD memory card (see Manual Data Logging Mode for
full setup instructions).
Automatic Data Logging: Setup the meter parameters in order to
automatically log data on a SD memory card. The number of data points
is limited by the size of the memory card.
Note: It is recommended to use a class 4 SDHC memory card between 1Gb
and 16Gb. Insert the SD memory card in the slot at the bottom of the meter.
The card must be inserted with the label side facing the rear of the meter.
Manual Data Logging Mode
In manual mode, the LOG button is pressed to manually log a reading on
the inserted SD card at any time. In order to setup the meter for manual
data logging, follow steps 1 through 7 below.
1. Set the data logging sampling rate to "0" seconds (see Setting the
Data Logging Sampling Rate section for details).
2. Press the REC button and "REC" will appear on the LCD.
3. Press the LOG button to enter manual data logging function as
indicated by "DATALOGGER".
4. The meter will also display P-X (X = memory position number
between 1 and 99).
5. Press the
and buttons to select one of the 99 data memory
positions in which to record.
6. Press the LOG button to save a reading to memory. 'DATALOGGER'
will ash each time a data point is stored.
Note: If a card is not inserted or the card is defective, the meter will ash
"EMPTY". In this case, power the meter OFF and try again with another
SD memory card or verify that the card is correctly inserted.
7. To exit manual data logging mode, press and hold the REC button
2 seconds to resume normal operation.