Thermocouple Temperature Measurement
1. Power on the instrument by pressing the Power Button
2. Plug a type K thermocouple probe in the Thermocouple Input
Socket
3. Select the Temperature function by pressing Function Button
4. Contact the Thermocouple Sensor Head with the object being
measured and the reading value will be displayed on the LCD
display
Humidity & Ambient Temperature Measurement
When the meter is applied in a new environment, it will take a few minutes
to reach a stable condition
1. Power on the instrument by pressing the Power Button
2. Select the Relative Humidity function by pressing the Function
Button
3. The reading value of relative humidity and temperature will be
displayed on the LCD display
Light Measurement
1. Power on the instrument by pressing the Power Button
2. Select the Light Measurement function by pressing the Function
Button until the light value is displayed
3. Press the Lux/Ft-cd Button to select measuring unit Lux or Ft-cd
Zero Offset Adjustment of Light Function
• For best results you should zero the light sensor prior to use in a dark
environment. To achieve this, place the light sensor end of the meter
under a desktop or at surface to block any light. Press the Unit/Zero
Button setting the meter indication to zero.
• The zero point can drift due to environment temperatures, battery
power, as well as various other reasons. It is recommended that the
zero be checked frequently using the above procedure.