User's Manual

SIM™ User Manual | Ref Code: SUM | GEN4_0024| Version: 8
181
Guidance and manufacturer’s declaration – electromagnetic immunity
The SIM™ system is intended for use in the electromagnetic environment specified below. The customer or the
user of the SIM™ system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601 TEST LEVEL
Compliance Level
Electromagnetic environment -
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
(>95% dip in
) for 0,5 cycle
40%
(60% dip in
) for 5 cycles
70%
(30% dip in
) for 25 cycle
<5%
(>95% dip in
) for 5 s
<5%
(>95% dip in
)
for 0,5 cycle
40%
(60% dip in
) for
5 cycles
70%
(30% dip in
) for
25 cycle
<5%
(>95% dip in
)
for 5 s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the SIM™
system requires continued operation
during power mains interruptions, it is
recommended that the SIM™ system be
powered from an uninterruptible power
supply or a battery.
Power Frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE
is the a.c. mains voltage prior to application of the test level.