7 Troubleshooting
7.1 General errors
7.1.1 Radio contact
7.1.2 USB stick
7.1.3 Software update
Problem Cause Solution
Parameters are not
transmitted.
The software is outdated.
Update the pipette
software.
Problem Cause Solution
No radio contact. The "Opt" operation
mode is not set.
Set the "Opt" operation
mode on the pipette.
The radio module is
deactivated.
Set the pipette radio
module to ON in the Wi-Fi
on/off menu.
The driver for the radio
module was not loaded.
Restart the control panel.
The pipette is always
offline.
Restart the control panel.
Restart the pipette.
Problem Cause Solution
USB stick is not detected. The USB stick is
incorrectly formatted.
Format the USB stick in
FAT32 format.
The USB stick is not
supported.
Use a different USB stick.
Problem Cause Solution
Software update is not
running.
The USB stick is not
detected.
Use a different USB stick.
Problem Cause Solution
Software update is not
running.
The necessary files are
missing.
Make sure that all files
are available.
Problem Cause Solution
Software update is not
running.
The software package has
been unpacked twice.
Unpack the software
package once only.