User's Manual

Table Of Contents
22
Operation
VisioNize
®
pipette manager
English (EN)
Qualified
5.3.3 Importing the liquid type
Prerequisites
A USB stick (file system FAT32) is available.
An import file is available on the USB stick.
1. Insert the USB stick.
2. Select the Menu > Liquid Manager menu.
3. Tap on the Import button.
4. Follow the instructions on the screen.
5.4 Using a single pipette
To be able to work with a pipette, it does not have to be assigned to a pipette set.
Prerequisites
A pipette with a radio module is available.
1. Call the home screen.
2. Open the Pipette Set drop-down list.
3. Select Select Single Pipette.
4. Select the desired pipette.
5.5 Combining pipettes to form a set
Pipettes can be assigned to a set in the Pipette Manager menu. A set can be selected on
the home screen and several pipettes are available at the same time.
Prerequisites
Pipettes are available for selection in the Manage Pipettes menu.
1. Call the Pipette Manager.
2. Tap on the New Set button.
3. Enter a name for the set and a description.