User`s manual
MegaPlus ES 1020 User’s Manual
Redlake MASD, Inc. Page 24 91000134-001 Rev. A
• VBIN 4/2: Vertical bin x 2 or x 4
Set to 1 to enable VBIN x 4
Set to 0 to enable VBIN x 2
• VBIN ENABLE
Set to 0 to enable normal operation
Set to 1 to enable vertical binning
• HBIN 4/2: Horizontal bin x 2 or x 4
Set to 1 to enable HBIN x 4
Set to 0 to enable HBIN x 2
• HBIN ENBLE
Set to 0 to enable normal operation
Set to 1 to enable horizontal binning
Exposure and Control Register 2
15 14 13 12 11 10 9 8
CR2_3 CROSS
HAIR
ENABLE
LONG_VBI
N
LVDS
CLOCK x 2
EXP BIT 11 EXP BIT 10 EXP BIT 9 EXP BIT 8
7 6 5 4 3 2 1 0
EXP BIT 7 EXP BIT 6 EXP BIT 5 EXP BIT 4 EXP BIT 3 EXP BIT 2 EXP BIT 1 EXP BIT 0
• EXP BIT 0 - 11
Manual Exposure
• LVDS CLOCK x 2 (note: set to 0 for normal operation, set to 1 for 2x LVDS clock)
Set to 0 to
Set to 1 to
• LONG V BIN
Set to 0 to disable V bin x 8 and 16
Set to 1 to enable V bin x 8 and 16 (Choose via VBIN x 4/2)
• CROSS HAIR ENABLE (note: use single tap mode for cross hair alignment)
Set to 0 to disable cross hair overlay
Set to 1 to enable cross hair overlay
• CR2_3