User`s manual

MegaPlus
®
Model ES 1.0
3.4.10 Transfer Pulse Delay
This command sets the time between the initial edge of the strobe pulse to the initial edge of the
transfer pulse in the triggered double exposure mode only.
TYPE IN: RESPONSE EXPLANATION
TPD xxx
CR-LF Sets the delay in increments of 1 microsecond. The
argument range is from 1 to 999 microseconds. If
TPD is set to zero, the EXE value controls the first
image exposure time permitting equal exposure
time for both images.
TPD?
Queries the current transfer pulse delay time
setting.
TPD xxx Means the time is set as indicated.
Sample Commands with Arguments:
TPD 0 The time set by the EXE command becomes the transfer pulse delay.
TPD 1 Sets the delay time to 1 microsecond.
TPD 999 Sets the delay time to 999 microseconds.
3.4.11 Transfer Pulse Width
This command sets the time of the transfer pulse in the triggered double expose mode only.
TYPE IN: RESPONSE EXPLANATION
TPW x
CR-LF Sets the delay in increments of one microsecond.
The argument range is from .2 to 5 microseconds.
TPW?
Queries the current transfer pulse width setting.
TPW x Means the time is set as indicated.
Sample Commands with Arguments:
TPW .2 Sets the delay time to .2 microseconds.
TPW 5 Sets the delay time to 5 microseconds.
91000064-004 Revision A 3-7 USER’S MANUAL