User`s manual
MegaPlus
®
Model ES 4.0/E
APPENDIX A – ERROR CODES AND DESCRIPTIONS
Code Name Description
ERROR_NONE
01 ERROR_SYNTAX Bad command
02 ERROR_XMIT UART error
03 ERROR_ARGUMENT Data out of range
04 ERROR_TYPE Data not type requested
05 ERROR_UNTERMINATED String not NULL terminated
06 ERROR_INDEX Can not INC or DEC command
07 ERROR_SPI SPI transmitting/receiving problem
08 ERROR_STATE Attempting to set camera to illegal state
09 ERROR_FLASH Flash read/write problem
10 ERROR_CATS CaTS send/receive problem
11 ERROR_DAYTONA DAYTONA command only
12 ERROR_CSUM Checksum error
13 Not Used
14 Not Used
15 ERROR_TEMP Camera temperature out of range
16 ERROR_SENS_TEC TEC out of range
17 ERROR_SENS_FAN Fan failure
18 Not Used
19 ERROR_P45 +45 volts failure
20 ERROR_P15 +15 volts failure
21 ERROR_P10 +10 volts failure
22 ERROR_P8 +8 volts failure
23 ERROR_P5 +5 volts failure
24 ERROR_N5 -5 volts failure
25 ERROR_N12 -12 volts failure
26 ERROR_FLASH_ACCESS Flash access denied
27 ERROR_FLASH_ERASE Flash erase operation failed
28 ERROR_FLASH_WRITE Flash write operation failed
29 ERROR_FLASH_READ Flash read operation failed
30 ERROR_FRBUFF_CONFIG_INIT Configuration initialization problem
31 ERROR_FRBUFF_CONFIG_TIMEOUT Configuration register problem
32 ERROR_FRBUFF_CONFIG Configuration data problem
33 ERROR_FRBUFF_CONFIG_DONE Configuration did not finish properly
34 ERROR_ANALOG_CONFIG_INIT Configuration initialization problem
35 ERROR_ANALOG_CONFIG_TIMEOUT Configuration register problem
36 ERROR_ANALOG1_CONFIG Configuration data problem
37 ERROR_ANALOG2_CONFIG Configuration data problem
38 ERROR_ANALOG_CONFIG_DONE Configuration did not finish properly
41 ERROR_PASSWORDS Password table did not load
42 WARNING_FAS Fast readout turned off
99 DOWNLOAD_OPERATION_OK For file download handshaking
91000116-001 Revision B I USER’S MANUAL