User manual

P a g e | 38
XXPRESS
USER MANUAL - 1.1
7.2.4 Running the Experiment
Before running an experiment the sample xxplate needs to be
prepared. All xxplates required to have a heat sealed film applied
to them to seal the contents and prevent evaporation during the
experiment. Details of how to seal the xxplates can be found in
the xxsealer manual.
NEVER PUT A XXPLATE IN THE XXPRESS SYSTEM WITHOUT
IT BEING HEAT SEALED. Only heat seal the xxplates with the
recommended heat sealing film.
NEVER PUT A LABEL ON THE BOTTOM OF THE XXPLATE.
Keep the bottom “black” surface of the xxplate clean and
never obstruct it by sticking a label on it as the unit measures
the temperature of the samples using the radiation emitted
from this surface.
To place the prepared and sealed xxplate into the xxpress
system touch the Front Door Release Button (See table 2) to open
the units loading door.
Figure 15