7.2.4 Running the Experiment
• Before running an experiment the sample xxplate needs to be
prepared. All xxplates required to have a heat sealed film applied
to them to seal the contents and prevent evaporation during the
experiment. Details of how to seal the xxplates can be found in
the xxsealer manual.
• NEVER PUT A XXPLATE IN THE XXPRESS SYSTEM WITHOUT
IT BEING HEAT SEALED. Only heat seal the xxplates with the
recommended heat sealing film.
• NEVER PUT A LABEL ON THE BOTTOM OF THE XXPLATE.
Keep the bottom “black” surface of the xxplate clean and
never obstruct it by sticking a label on it as the unit measures
the temperature of the samples using the radiation emitted
from this surface.
• To place the prepared and sealed xxplate into the xxpress
system touch the Front Door Release Button (See table 2) to open
the units loading door.