Operating instructions

07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 20-3
20. Quick Reference
WARNING
The solution bag must be positioned properly on the heater pan,
covering the silver heater sensor button. Failure to properly position
the solution bag can result in overheated or underheated dialysis
fluid.
WARNING
DO NOT use external heating sources (i.e., microwave oven) to warm
solution bags. This can result in overheated solution delivered into
your peritoneal cavity.
2. Place a solution bag on the
heater pan.
3. Turn on the cycler.
Press the On/Off switch to the
ON position.
Observe that the characters on
the display screen turn on and
then off for several seconds.
The current mode (STANDARD
MODE or LOW FILL MODE)
appears for a few seconds.
When the system is ready,
PRESS GO TO START appears.