19. Technical Data
19-6 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Table 19-2. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The HomeChoice/HomeChoice PRO APD System is intended for use in the electromagnetic environment specified below.
The user of the system should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level Compliance Level Electromagnetic Environment – Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
hospitals or buildings used for commercial or domestic
purposes.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
hospitals or buildings used for commercial or domestic
purposes.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
) for 0.5 cycles
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 seconds
<5% U
T
(>95% dip in
U
T
) for 0.5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 seconds
Mains power quality should be that of a typical public
low-voltage power supply network that supplies
hospitals or buildings used for commercial or domestic
purposes.
If the user of the system requires continued operation
during power mains interruptions, it is recommended
that the system be powered from an uninterruptible
power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels
characteristic of a typical public low-voltage power
supply network that supplies hospitals or buildings used
for commercial or domestic purposes.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.