Operating instructions
07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-67
18. Troubleshooting
18.10.2 Return to Therapy After an Emergency Disconnect
WARNING
Follow aseptic technique taught by your dialysis center
when handling lines and solution bags to reduce the
possibility of infection. Always put on a face mask and wash
and dry (or disinfect) your hands thoroughly.
Steps to return to therapy
1. Using aseptic technique, prepare to connect:
a. Remove the FlexiCap disconnect cap from the patient line connector.
b. Remove the MiniCap disconnect cap from the transfer set connector.
c. Remove the patient line connector from the organizer.
d. Connect the patient line connector to the transfer set.
e. Open the clamp on the patient line.
f. Open your transfer set.
g. Press
GO.
h. Discard the used MiniCap disconnect cap and FlexiCap disconnect cap.
2. Your therapy continues.