07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-21
18. Troubleshooting
18.3.10 Warming Solution
WARMING SOLUTION
Display Message:
Cause: The temperature of the fluid in the heater bag as measured
by the cycler is below 33°C.
Fluid delivery is not allowed at this time.
The message continues to appear while the fluid in the
heater bag is being heated.
The message clears 5 (five) minutes after the heater bag
temperature reaches 33°C.
The message can also appear for an extended period if the
heater bag is not in full contact with the heater plate during
normal operation.
To Correct: If the alarm becomes a WARMING TIMEOUT alarm after
45 minutes:
1. Press
STOP to mute the alarm.
2. Press
GO to gain another 45 minutes to heat the
solution.
3. Press the heater bag down against the heater plate to
improve the heating rate.
A SYSTEM ERROR appears if the heating system fails.
WARMING SOLUTION