Safety Instructions
18
Note Possible Cause Action
One garment inflates but the second
one does not.
The second garment does not receive air. Examine its hoses for kinks, twists and
folds.
The pump stops working, the Status
Indicator turns On in yellow.
Hose is not connected properly to
garment or pump, or prong plug is
not inserted into unused air outlet.
Examine and fasten all air connections.
When treating only one limb, always
plug the unused air outlet with the prong
plug provided with the pump. If all air
connections are OK and the problem
persists, contact Therabody.
The pump works at a very low pressure,
regardless of the pressure set by the user.
Defective garment Replace garment and check again.