User's Manual

USER MANUAL P/N UM1400-00-09 12/30/19 Pg 34
Recovery Force 10022 Lantern Rd., Suite 100, Fishers, IN 46037 www.recoveryforceusa.com
012: Board Over-Temperature
013: Strap Read/Write Error
014: RTC Error
A valid Strap is not connected to
the Controller.
Connect the Controller to a MAC System
Strap.
The battery is critically low.
Replace the battery with a fully charged
battery from the Charging Hub.
Date or time of Controller does
not match the date or time of
the data in the Strap from the
previously used Controller.
The device is unable to keep the previous
data on the strap due to an incorrect
date/time. Press OK to continue because the
previous data is unable to be retrieved.
Strap is too loose on patient’s
leg.
Adjust and tighten Strap on patient’s leg.
Strap is too tight on patient’s leg.
Adjust and loosen Strap on patient’s leg.
Strap has reached its maximum
usage.
Initiate a new MAC System Strap.
A nonauthorized Strap is
attempting to be used.
Initiate new authentic MAC System Strap